β MSO5000 Live Monitor β Power Analysis Setup Checklist AC
π For Grid Power Measurement with Differential Current Probe (High-Voltage AC)
π§° Hardware Assumptions
- Voltage probe: CH1 β measures 230V AC (grid voltage)
- Current probe: CH4 β differential current probe with attenuation (e.g. 5Γ), placed around load/live wire
- Scope type: RIGOL MSO5000
- Load type: AC (resistive or reactive), 50Hz
- Direction: Current flows into the load (positive power)
π Oscilloscope Setup (on RIGOL MSO5000)
CH1 β Grid Voltage
- Connect across LβN (line to neutral)
- Set unit:
VOLT
- Set probe:
1X
(or 10X depending on your voltage probe) - Enable 20MHz BW Limit β (filters noise, improves PF accuracy)
CH4 β Grid Current (Differential Probe)
- Connect current probe around line wire
- Set unit:
AMP
β (crucial!) - Set probe:
5X
(match your differential probe attenuation) - Enable Invert β (if current direction is flipped β makes power positive)
- Enable 20MHz BW Limit β (stabilizes power factor and FFT phase)
βοΈ Live Monitor: Power Analysis Tab Setup
Field | Value / Action | Why |
---|---|---|
Voltage Ch |
1 |
CH1 measures grid voltage |
Current Ch |
4 |
CH4 measures current |
Correction Factor |
1.0 |
No scaling needed β scope already shows real amps |
Current Probe Type |
Clamp β
|
Internally handled as unit = AMP |
Probe Value |
1.0 |
β οΈ Leave at 1.0 β no further scaling |
DC Offset |
β Usually off | Optional β turn on if baseline is offset |
25M [v]/[i] |
β Off | Not needed for 50Hz AC |
β€ β‘ Measure |
Click to start | Done β |
π§Ύ Recommended Checks
- β Vrms should be around 230V
- β Irms should match what clamp/differential probe shows
- β Real Power (W) should be positive (if direction is correct)
- β Apparent Power (VA) = Vrms Γ Irms
- β Power Factor matches scope (~0.04 for small reactive loads)
- Use
π Plot Last
to view PQ triangle with S/P/Q vectors
β οΈ Notes on Accuracy
Case | What to Do |
---|---|
Current reads negative | β
Enable Invert on CH4 |
Power is 10Γ too low | β
Check probe setting: 5X vs 1X |
PF is unstable | β
Enable BW Limit on CH1 & CH4 |
Scope reads wrong amps | β
Make sure CH4 unit = AMP |
β Why This Setup βJust Worksβ
Because CH4 is set to AMP and the scope knows the probe attenuation (
5X
), both the Rigol and Live Monitor treat the data as real-world Amps, no manual scaling needed.
Inversion, probe setting, and unit type are all respected by both systems β so the numbers match perfectly.
π Quick Reference Table
Signal | CH | Scope Unit | Probe Setting | GUI Probe Type | GUI Probe Value |
---|---|---|---|---|---|
Grid Voltage | 1 | VOLT |
1X or 10X |
β | β |
Grid Current | 4 | AMP |
5X |
Clamp |
1.0 |
π§ If Rigol PQ and Live Monitor match exactly β you're doing it right.