Skip to content

βœ… MSO5000 Live Monitor β€” Power Analysis Setup Checklist DC

πŸ§ͺ For Shunt-Based DC Power Measurements (Low-Side Current Sensing)


🧰 Hardware Assumptions

  • DC power supply (e.g., 40V / 3.5A)
  • Shunt resistor on the negative rail (e.g., R010 = 0.01Ξ©)
  • Voltage measurement: CH2
  • Shunt voltage measurement: CH3

πŸ›  Oscilloscope Setup (on RIGOL MSO5000)

CH2 β€” Voltage Channel

  • Connect to the positive side of your load or power supply
  • Set unit: VOLT
  • Set probe: 10X (or 1X, depending on your probe type)
  • Enable 20β€―MHz BW Limit βœ…

CH3 β€” Shunt Voltage (Current)

  • Connect across the shunt resistor (GND/negative side)
  • Set unit: VOLT ⚠️ (not AMP!)
  • Set probe: 1X (unless using attenuated probe or amp clamp)
  • Enable Invert if waveform is upside down
  • Enable 20β€―MHz BW Limit βœ…

βš™οΈ Live Monitor: Power Analysis Tab Setup

Field Value / Action Why
Voltage Ch 2 CH2 measures supply/load voltage
Current Ch 3 CH3 connected across shunt
Correction Factor 1.0 Leave as-is unless doing manual calibration
Current Probe Type Shunt βœ… Required for converting volts β†’ amps
Probe Value (Ξ©) 0.01 for R010 Your shunt's resistance
Expected P (W) (Optional) e.g. 250 β†’ hit βš™ Cal Auto-tune if you know expected power
DC Offset βœ… Check if signal has DC baseline drift Improves PF and FFT accuracy
25M [v]/[i] ❌ Leave off unless needed Only enable for high-resolution waveform fetches
➀ ⚑ Measure Click to run power analysis You’re done βœ…

  • βœ… Does Real Power β‰ˆ Voltage Γ— Current from your DC supply?
  • βœ… Is Irms ~ what your shunt should show?
  • βœ… Does PF or phase angle make sense for your load?

Use πŸ“ˆ Plot Last to visualize your PQ triangle.


⚠️ Important: Rigol PQ Will Be Wrong

Rigol’s Power Quality feature assumes CH3 is already in Amps.
If you're measuring shunt voltage, Rigol will report 100Γ— too little power.

πŸ“Œ Only Live Monitor properly scales shunt measurements.


🏁 Quick Reference Table

Signal CH Scope Unit GUI Probe Type Probe Value
Voltage (V) 2 VOLT β€” β€”
Shunt Voltage 3 VOLT Shunt 0.01
Current Clamp 3 VOLT Clamp e.g. 1000
Current Direct 3 AMP Clamp e.g. 1.0

🧠 Still confused? If Rigol says β€œ2.5W” and Live Monitor says β€œ250W” β€” the latter is correct.