Skip to content

Understanding Current Probe vs Differential Probe (Across Shunt) in Oscilloscope Power Measurements

Scenario

You are measuring current in a circuit using an oscilloscope.
You have two options:

  1. Current Probe (e.g., clamp, Hall effect, Rogowski)
  2. Differential Probe across a shunt resistor

Configuring your oscilloscope or analysis software as if your diff probe was a “current probe” (unit: AMP) gave incorrect results. Why?


Key Differences

Probe Type Senses Output Voltage Means Software Conversion Needed? Example Scaling
Current Probe Magnetic field Proportional to current No (if scaling correct) 1V/A, 10mV/A, etc
Diff Probe+Shunt Voltage difference Voltage across shunt Yes, V/Ω

How They Work

1. Current Probe (Clamp, Hall, Rogowski)

  • Senses: Magnetic field around the conductor (no circuit interruption)
  • Outputs: Voltage directly proportional to the measured current
    • e.g., 1V out = 1A flowing (if probe designed for this scaling)
  • Oscilloscope Setup: Set channel unit to AMP, enter correct probe scaling.
  • Software: No conversion needed if probe scaling is set—voltage is already "current".

2. Differential Probe (Across Shunt Resistor)

  • Senses: Voltage drop across the shunt resistor.
  • Outputs: Voltage proportional to current, but must use Ohm's law:
    • I = V / R_shunt
    • e.g., 10mV across 0.01Ω shunt = 1A
  • Oscilloscope Setup: Channel unit should be VOLT (not AMP).
  • Software: Enter shunt resistance (Ω) so software converts voltage to current.

Common Pitfall

Setting your oscilloscope channel to “AMP” when using a diff probe + shunt does NOT make the measurement correct.
- The scope only changes the displayed unit, not the actual conversion. - Unless the scope’s probe settings also know your shunt value, results will be incorrect.


Correct Setup for Differential Probe + Shunt

  1. Scope channel unit: VOLT
  2. In analysis software:
    • Select Probe Type: Shunt
    • Enter shunt value in Ω
    • The software will compute current as I = V / R_shunt

Summary Table

Setup Scope Unit Probe Type Value Software Conversion Correct?
Clamp Current Probe AMP Clamp (V/A) None (if scaling set) Yes
Diff Probe Across Shunt VOLT Shunt (Ω) I = V / R_shunt Yes
Diff Probe + Shunt, unit: AMP AMP Clamp 1.0 None No

Practical Advice

  • Use current probe + “AMP” only if you have a real current probe.
  • For shunt+diff probe, always use VOLT and set the correct shunt value in your analysis software.
  • Double-check your software scaling and units!

If you see weird current readings or power calculations, check your probe type, scaling, and software settings.